Authorised Representative for a Medical Device Manufacturer in Russia and the EAEU

An authorised representative (also written "authorized representative") is a legal entity or individual entrepreneur resident in an EAEU member state that a manufacturer from outside the Union appoints to represent it before the regulators and to carry responsibility for the medical device on the Union market. Any manufacturer based in China, the European Union, the United States, Turkey, Korea or any other third country needs one — the requirement is mandatory, not optional, and the authorised representative is normally the only party that can act as applicant in the registration procedure. Appointing one does not by itself take long: the appointment is a contractual matter and can be completed in days. The registration procedure it enables takes about five months of regulator time and typically 12 to 24 months as a project.

This page explains what the role legally is, what it obliges you to do, how it is appointed and maintained, and what happens when it lapses.

What is an authorised representative under EAEU law?

The definition comes from the registration Rules approved by EAEU Council Decision No. 46 of 12 February 2016. Three elements of that definition matter, and each has practical consequences:

  • It must be a legal entity or an individual entrepreneur. An informal arrangement, a distributor's verbal commitment or a branch without legal personality does not satisfy the definition.
  • It must be resident in a member state of the Union — Russia, Kazakhstan, Belarus, Armenia or Kyrgyzstan. The representative is the point at which the manufacturer becomes reachable under Union and national law.
  • It must be authorised by the manufacturer under a contract or another document provided for by the legislation of the member states, to represent the manufacturer's interests and to bear responsibility in relation to the circulation of the device within the Union and compliance with the mandatory requirements for it.

That last element is the one manufacturers most often underestimate. The representative is not a mailbox. The rules place responsibility for compliance with the General Requirements for safety and efficiency — approved by EAEU Council Decision No. 27 of 12 February 2016 — jointly on the manufacturer and its authorised representative. Signing the application is a declaration with legal consequences.

When is an authorised representative mandatory — and when is it not?

The dividing line is where the manufacturer is established, not where the product is sold.

| Where the manufacturer is established | Is an authorised representative required? | What follows |

|---|---|---|

| In an EAEU member state (legal entity or individual entrepreneur) | Not required by Union law | The manufacturer may act as applicant itself; no representative details are entered |

| In another EAEU member state | Not required if the manufacturer applies directly | The manufacturer may represent itself before the reference state |

| Outside the Union — China, EU, US, Turkey, Korea and elsewhere | Required | Appointed by contract, acting throughout the period of circulation, with its details kept up to date in the registration dossier |

Even where the role is not strictly required, a manufacturer based abroad often needs a local party for operational reasons: to receive notifications and queries, correspond with the expert organisation, and respond within statutory deadlines. That is project practicality rather than a regulatory requirement, and we do not present one as the other.

Which documents define the role?

EAEU Council Decision No. 46 of 12 February 2016

  • What it governs in relation to this role: the definition of the manufacturer's authorised representative; the obligation on manufacturers from third states to appoint one and to keep the information about it up to date; the rule that responsibility for compliance with the General Requirements is borne by the manufacturer and its authorised representative; the requirement to include a copy of the document confirming the representative's powers in the registration dossier; the application forms on which the representative appears as applicant; and the procedure for changing the representative through the notification route.
  • Primary source: docs.eaeunion.org

EAEU Council Decision No. 27 of 12 February 2016

  • What it governs: the General Requirements for the safety and efficiency of medical devices, and the requirements for their labelling and operational documentation. These are the requirements for which the authorised representative shares responsibility.
  • Primary source: docs.eaeunion.org

EAEU Council Decision No. 106 of 10 November 2017

  • What it governs: the requirements for implementing, maintaining and evaluating a quality management system for medical devices according to potential risk. The obligation sits on the manufacturer, but the authorised representative is the party through which it is verified in practice, including during manufacturing inspection.
  • Primary source: docs.eaeunion.org

EAEU Board Decision No. 174 of 22 December 2015

  • What it governs: the Rules for monitoring the safety, quality and efficiency of medical devices. Post-market obligations continue for as long as the device is on the market, and the authorised representative is the standing contact point for them.
  • Primary source: docs.eaeunion.org

Federal Law No. 323-FZ of 21 November 2011

  • What it governs: the fundamentals of health protection in Russia. Article 38 governs the circulation of medical devices and admits to the market only devices that have been registered; Article 96 provides for monitoring of the safety of medical devices.
  • Primary source: consultant.ru

Government Decree No. 1684 of 30 November 2024

  • What it governs: the Russian national Rules of state registration of medical devices, which operate in parallel with Union law during the transition period. These Rules apply until 31 December 2028.
  • Primary source: consultant.ru

What does an authorised representative actually do?

The role is broader than signing applications. In the projects we run it consists of eight blocks of work.

1. Representing the manufacturer before the regulator. Filing the applications for registration and examination, dealing with the competent authority and the expert organisation of the reference state, receiving notifications, queries and decisions.

2. Answering queries from the examination. Where the expert organisation needs more material, it sends a query once. The applicant has up to 60 working days to respond; miss that, and the decision is taken on the documents the regulator already holds.

3. Ensuring the device complies with mandatory requirements. Responsibility for compliance with the General Requirements is shared between the manufacturer and the representative.

4. Handling changes to the registration dossier. Changes are made either through examination or through the notification procedure, and in every case the application is filed by the applicant — the manufacturer or its authorised representative.

5. Keeping the dossier information current. The Rules require the information about the representative contained in the registration dossier to be maintained in an up-to-date state. A change of address, a reorganisation or a change of representative is a reason to start a procedure, not to postpone one.

6. Renewing time-limited documents. Contracts, powers of attorney and supporting documents have validity periods. Renewing them before expiry is part of the role.

7. Safety monitoring and complaint handling. Adverse event reports, complaints and corrective actions all converge on the representative as the party within the regulator's jurisdiction.

8. Taking part in manufacturing inspections and state control. The representative is the point of contact for inspections and for document requests from supervisory authorities.

Can a foreign manufacturer register a device without one?

No — and the way the rule bites is procedural rather than punitive, which is what makes it easy to underestimate.

  • The Rules require manufacturers from third states to appoint an authorised representative and to maintain current information about that representative in the registration dossier.
  • The applicant is the manufacturer or its authorised representative. There is no third category. Since Order No. 750 of Roszdravnadzor came into force on 27 September 2026, even the developer of a device is no longer an independent applicant under the Russian national procedure.
  • The dossier must contain a copy of the document confirming the representative's powers.
  • The completeness check takes 7 working days. If the dossier is incomplete, the applicant is notified and given up to 30 working days to remedy the position. If that period passes without a cure, the application and documents are returned within 3 working days.

The consequence is that a missing or defective document on representation is not a minor remark. It is a ground for returning the entire application, and a return means re-entering the procedure with a new clock. We have seen projects lose a quarter to this alone.

How do you appoint an authorised representative?

1. Check the candidate. The representative must be resident in an EAEU member state as a legal entity or an individual entrepreneur, with registration documents to prove legal capacity.

2. Conclude a contract with the manufacturer. The contract or another document provided for by member-state legislation is the basis of the powers. It should spell out: representing the manufacturer before competent authorities and expert organisations; signing applications; conducting correspondence; access to the registration dossier; handling changes; assisting with manufacturing inspection and state control; and the procedure for exchanging safety information.

3. Execute the document confirming the powers. A copy of that document goes into the registration dossier, and documents in a foreign language must be accompanied by a certified translation into Russian.

4. Synchronise it with the rest of the dossier. The name and address of the representative must match everywhere it appears: in the applications, in the forms, in the labelling artwork, and in the instructions for use where the representative is named.

5. Enter and maintain the information. If the device is already registered and the representative changes, the change is processed through the registration dossier.

How long does each step take?

| Step | Period | Notes |

|---|---|---|

| Contract and document confirming powers | days, not months | A contractual matter between manufacturer and representative |

| Completeness check of the dossier | 7 working days | If the powers document is missing, the application is returned |

| Examination in the reference state | 45 working days | The substantive stage of registration |

| Review by each recognition state | 30 working days | From a date defined in the Rules |

| Registration decision and issue of the certificate | 10 + 10 working days | The certificate is issued without expiry |

| Change of representative, notification procedure | 15 working days | From the decision to start the procedure; no recognition-state agreement needed |

How is the authorised representative changed after registration?

Changing the representative is one of the cases expressly covered by the notification procedure for amending the registration dossier. That procedure applies where the changes concern information about the manufacturer or its authorised representative — for a legal entity, its name and the address of its location or place of business; for an individual entrepreneur, their name, address and identity document details.

Under the notification route the authority checks the completeness and accuracy of the submitted information, and the change itself is made within 15 working days of the decision to start the procedure. The notification procedure does not require agreement from the recognition states, and recognition states are not named in the application. Where the change affects the registration certificate, a new certificate is issued bearing the previous number and the date of the change.

Two practical warnings. First, the outgoing and incoming representatives must both be properly documented — an unclear handover is a common cause of returned applications. Second, because the same name and address appear in labelling and in the instructions for use, a change of representative usually triggers consequential artwork and documentation updates. Planning those alongside the dossier change avoids a second cycle.

What happens if the representative's details are out of date?

There is no need to invent penalties. The consequences follow directly from the rules, and they are serious enough on their own:

  • The device cannot be registered. Without a valid document confirming the representative's powers, the completeness check fails.
  • Placing on the market is blocked. Registration is a mandatory condition for releasing a device into circulation in the member states. No registration means no lawful import for sale — the device may be bought and stored, but it cannot be supplied.
  • Routine procedures stall. A change of manufacturer address, a new model designation, an updated standard or the renewal of an expiring document are all filed by the applicant. Without an active representative, nobody can file.
  • Supervision becomes difficult. State control over the circulation of medical devices presupposes a party in the country answerable for the device. A registry entry that no longer matches reality is an obvious discrepancy.
  • Complaints and adverse events lose their channel. They arrive at the address in the Union's jurisdiction — or they do not arrive at all.

In our experience the most expensive consequence is not a fine. It is stopped shipments: the goods are purchased and sitting in a warehouse, and they cannot be imported for sale until the chain of manufacturer, representative and dossier is restored.

Why manufacturers appoint KREDO.TECH

We work with foreign medical device manufacturers, including manufacturers from China, and we act as authorised representative as part of a full registration cycle. In practice that means:

  • a contract and supporting documents prepared so the dossier passes the completeness check first time;
  • named specialists who answer correspondence with the expert organisation and the competent authority, rather than a generic inbox;
  • query responses drafted with the technical team, so the single query the Rules allow is answered fully;
  • changes to the registration dossier handled as they arise, including changes of representative, address and model range;
  • safety information, complaints and inspection requests routed to one responsible party;
  • the same party accountable from the first application through to post-market obligations, so nothing falls between a consultant and a distributor.

Our services are listed on the services page and indicative pricing is on the tariffs page. The specialists who would be named in your contract are described on the team page. If your device is an in vitro diagnostic, start with the IVD registration page; for the procedure itself, see EAEU registration.

What to do next

Send us the manufacturer's details, the device's intended purpose and current classification, and any registration history in Russia or the EAEU. We will confirm whether an authorised representative is required in your case, set out exactly what the contract and the supporting documents must contain, and give you a timeline from appointment to registration.

There is no charge for that first assessment and it usually takes one working day. To start, use the contacts page.

Frequently asked questions

Who can act as an authorised representative?
A legal entity or an individual entrepreneur resident in an EAEU member state — Russia, Kazakhstan, Belarus, Armenia or Kyrgyzstan. An informal arrangement, a distributor's verbal commitment or a branch without legal personality does not satisfy the definition.
Is an authorised representative mandatory for a manufacturer from outside the EAEU?
Yes. A manufacturer based in China, the European Union, the United States, Turkey, Korea or any other third country must appoint one, and the authorised representative is normally the only party that can act as applicant in the registration procedure.
How long do the appointment and the registration take?
The appointment itself is a contractual matter and can be completed in days. The registration procedure it enables takes about five months of regulator time and typically 12 to 24 months as a project.
Who is responsible for compliance with the General Requirements?
Responsibility is borne jointly by the manufacturer and its authorised representative, under the General Requirements approved by EAEU Council Decision No. 27 of 12 February 2016. The representative is not a mailbox — signing the application is a declaration with legal consequences.
Which documents define the role and how is the representative changed?
The definition, the duty to appoint a representative and the procedure for changing it come from EAEU Council Decision No. 46 of 12 February 2016, and a copy of the document confirming the representative's powers goes into the registration dossier. A change is made through the notification route.