Medical Device Registration in the EAEU: One Procedure, One Certificate, Several Markets

Medical device registration in the EAEU is a single procedure in which the competent authority of one member state — the reference state — issues a registration certificate that is valid without expiry in every member state named on it. It is mandatory for any manufacturer that wants to place a device on the market in Russia, Kazakhstan, Belarus, Armenia or Kyrgyzstan, whether the device is made inside the Union or imported from China, the European Union, the United States, Turkey or Korea. The administrative part of the procedure is short on paper: 7 working days to check that the dossier is complete, 45 working days for the examination, 30 working days for each recognition state, then 10 plus 10 working days to take the registration decision and issue the certificate. That is roughly five months of regulator time. The full project — with testing, translation and dossier preparation — normally runs 12 to 24 months.

This page explains what the procedure actually requires, which documents govern it, where the time goes, and what we will not promise you.

What EAEU medical device registration is — and what it replaces

Registration under Union law replaces registering a device separately in each country: one application, one examination, one certificate. The certificate confirms that the device may be released into circulation in the member states listed in it, and it does not expire.

Three consequences follow from that design, and they shape every project:

  • You choose one reference state. For a single device you may not select two or more reference states. The reference state's authority and expert organisation carry out the examination and issue the certificate.
  • You choose at least one recognition state. Recognition states review the expert conclusion produced by the reference state and either confirm or refuse to confirm it. If a recognition state does not confirm, the information about that state is removed from the certificate — the device is registered for the remaining states only.
  • Requirements are identical for domestic and imported devices. A device manufactured inside the customs territory of the Union and a device imported from a third country are assessed against the same rules.

One further requirement applies to manufacturers outside the Union: a manufacturer from a third state must appoint an authorised representative resident in a member state and keep its details up to date in the dossier. That role is covered on the authorised representative services.

When must you register under EAEU rules — and when can you still use national rules?

This is the first commercial question in any 2026 project, and it changes the budget.

Registration under Union law was originally meant to become the only route from 1 January 2026. Instead, on 29 December 2025 the member states signed a Protocol extending the transition period: applications may be submitted until 31 December 2027 either under Union law or under the national legislation of a member state. Russia ratified the Protocol by Federal Law No. 288-FZ of 4 August 2026.

In practice the choice looks like this:

  • National registration (Roszdravnadzor). A well-established route with accumulated practice. The certificate is valid in one country only. If you later want Kazakhstan, Belarus, Armenia or Kyrgyzstan, the device must be registered again there.
  • EAEU registration. One procedure and one certificate covering several states. The evidence requirements are broader and the dossier is built to Union rules rather than to Russian ones.

Our working conclusion: decide by sales horizon. If the device is planned for Russia only and time to market is critical, the national route remains faster today. If two or more member states are in scope, one Union procedure is normally cheaper than two or three national ones.

Which documents govern EAEU registration?

The procedure is set by acts of the Eurasian Economic Commission. Numbers and dates below were verified against the EAEU legal portal, and every card links to the primary source — this is a field where secondary summaries drift.

EAEU Council Decision No. 46 of 12 February 2016

  • What it governs: the Rules of Registration and Examination of Safety, Quality and Efficiency of Medical Devices — the core procedural act. It defines the reference state and recognition states, the registration dossier, the timelines for each stage, changes to the dossier, duplicate certificates, suspension and cancellation. It also defines key terms, including "authorised representative of the manufacturer".
  • Primary source: docs.eaeunion.org

EAEU Council Decision No. 27 of 12 February 2016

  • What it governs: the General Requirements for Safety and Efficiency of Medical Devices, and the requirements for their labelling and operational documentation.
  • Primary source: docs.eaeunion.org

EAEU Council Decision No. 28 of 12 February 2016

  • What it governs: the Rules for technical testing of medical devices.
  • Primary source: docs.eaeunion.org

EAEU Council Decision No. 29 of 12 February 2016

  • What it governs: the Rules for clinical and clinical-laboratory testing (studies) of medical devices.
  • Primary source: docs.eaeunion.org

EAEU Council Decision No. 38 of 16 May 2016

  • What it governs: the Rules for studies (tests) to assess the biological action of medical devices.
  • Primary source: docs.eaeunion.org

EAEU Council Decision No. 106 of 10 November 2017

  • What it governs: the requirements for implementing, maintaining and evaluating a quality management system for medical devices, depending on the potential risk of their use.
  • Primary source: docs.eaeunion.org

EAEU Board Decision No. 173 of 22 December 2015

  • What it governs: the Rules for classifying medical devices according to the potential risk of use. Classes 1, 2a, 2b and 3 apply to general devices; IVDs have their own scheme within the same act.
  • Primary source: docs.eaeunion.org

EAEU Board Decision No. 174 of 22 December 2015

  • What it governs: the Rules for monitoring the safety, quality and efficiency of medical devices — the post-market obligations that continue after registration.
  • Primary source: docs.eaeunion.org

EAEU Board Decision No. 177 of 29 December 2015

  • What it governs: the Rules for maintaining the EAEU nomenclature of medical devices. Every device must be assigned a type within that nomenclature.
  • Primary source: docs.eaeunion.org

Who does what: reference state and recognition states

The applicant is the manufacturer or its authorised representative. There are no other categories of applicant.

The reference state's authority and expert organisation examine the device, prepare the expert conclusion and take the registration decision. Recognition states review that conclusion and send back confirmation or a reasoned refusal. A refusal by one recognition state is not fatal — it results in that state being excluded from the certificate.

When choosing a reference state, the fee is only one factor. Administrative practice and queue depth differ noticeably between member states, and over a project of this length those differences usually outweigh the fee.

The procedure step by step, with the statutory deadlines

The table below reflects the periods set in Decision No. 46. They apply to the regulator and the expert organisation; time the applicant takes to respond to queries is not counted within them.

| Stage | Statutory period | What it means in practice |

|---|---|---|

| Check of completeness and accuracy of the dossier | 7 working days | If something is missing or inaccurate, you are notified |

| Time to remedy identified defects | up to 30 working days | If you do not respond, the application is returned within 3 working days |

| Decision to start the registration and examination procedure | 3 working days | The dossier is placed in the information system |

| Examination of the device and preparation of the expert conclusion | 45 working days | The substantive stage: classification, nomenclature, technical characteristics, labelling, production data |

| Inspection of the manufacturing site, where required | up to 90 working days in aggregate | Organised in accordance with the quality management system requirements |

| Response to a query from the examination | up to 60 working days | Queries are sent once; if no answer comes, the decision is taken on the documents available |

| Review of the expert conclusion by each recognition state | 30 working days | Counted from a date defined in the Rules |

| Registration decision | 10 working days | Taken once all recognition states confirm, or their period expires |

| Issue of the certificate | 10 working days | From the date of the registration decision |

Three details cause most delays. The query response period — up to 60 working days — sits outside the official clock, so one poorly answered query can add three months without any deadline being breached. An inspection adds up to 90 working days, also excluded from the examination period. Queries are sent only once, so an incomplete answer is not followed by a second chance.

What goes into the registration dossier

The dossier is a structured set of documents defined in the Rules. In substance it contains:

  • the applications for registration and examination, in the established forms;
  • the document confirming the powers of the manufacturer's authorised representative, where one is appointed;
  • technical and operational documentation for the device, including technical specifications;
  • results of technical testing, biological action assessment, and clinical or clinical-laboratory testing — or first-party evidence already held by the manufacturer;
  • labelling, packaging and instructions for use;
  • information about manufacturing sites and the quality management system, including manufacturing inspection reports where an inspection was carried out;
  • information about adverse events, recalls and safety notices, together with a post-market safety and performance data collection plan.

Two dossier rules are absolute. Documents are submitted in Russian and electronically; where paper is required, electronic copies must also be supplied as PDF files with a searchable text layer. Any foreign-language document needs a certified translation into Russian.

A recurring and expensive misconception is that a dossier prepared for national registration will "almost do" for the EAEU. The structure looks similar; the content does not transfer. Documents written to Russian national rules cite different legislation, use different terminology and follow a different logic of conformity. Two documents in particular must be synchronised with everything else: technical specifications for the device and the clinical report.

Which Russian deadlines changed in 2026?

Because most of our clients also look at the Russian national route, one clarification matters. The EAEU procedure is governed by Decision No. 46. The national Russian procedure is governed by separate rules approved by Government Decree No. 1684 of 30 November 2024, and by the administrative regulation of Roszdravnadzor.

That regulation is Roszdravnadzor Order No. 750 of 19 August 2026, registered with the Ministry of Justice on 15 September 2026 under No. 88279 and in force since 27 September 2026. It replaces Order No. 3371 of 6 May 2019 and applies until 31 December 2028. It sets the following maximum periods for the state service, counted from registration of the application and documents and measured in working days:

| Application | Maximum period |

|---|---|

| Registration of a device whose efficacy must be confirmed by clinical trials with human participation | 50 working days |

| Registration of a device not requiring such clinical trials | 31 working days |

| Registration of a domestically produced device | 10 working days |

| Changes to the registration dossier not requiring examination | 18 working days |

| Changes to the registration dossier requiring examination | 36 working days |

| Changes concerning medical software using artificial intelligence | 10 working days |

| Cancellation of state registration | 7 working days |

Two further changes matter. The developer of a device is no longer an independent applicant — only the manufacturer or its authorised representative may apply. And the result is now a registry entry in the State Register rather than a separately issued document.

We do not promise faster than the regulation allows. Anyone quoting a two-month national registration for a device requiring clinical trials is quoting a period the regulation does not provide for. For devices covered by the IVD requirements, the clinical evidence stage is usually the long pole, and no administrative deadline changes that.

How long does EAEU registration really take?

Separate the statutory clock from the project.

The statutory clock. Seven working days to check completeness, 45 for the examination, 30 for recognition, 10 and 10 for the decision and the certificate — about 102 working days, roughly five months, assuming no queries and no inspection.

The project. Before the clock starts you must assemble the evidence. Depending on the class, that means technical testing, biological action assessment, clinical or clinical-laboratory testing, a quality management system assessment, translation and certification, and a dossier built to Union requirements. In our experience a first-time EAEU registration of a class 2b device, starting from documentation prepared for another market, takes 12 to 24 months to a certificate.

Where projects lose months

1. The device is classified by one rule only. The rules require you to assess every relevant provision and take the highest resulting risk class. Devices that fall under several provisions are routinely under-classified at the planning stage.

2. The intended purpose is too broad. Broad wording invites a higher class and a larger testing programme; a specific intended purpose and target population reduce the evidence burden legitimately.

3. The dossier is a translation exercise. A translated dossier that still argues from foreign legislation will generate a query at the examination stage — when the clock is already running.

4. Manufacturing inspection is left until requested. The inspection compares what the dossier says about production with how production actually works; closing that gap beforehand is materially cheaper.

5. Complaints and adverse-event data are not gathered. The dossier must describe known adverse events, recalls and safety notices. Empty sections are not read as "no problems".

6. The authorised representative is appointed late. Without a valid document confirming the representative's powers, the completeness check fails and the whole application is returned.

What to do next

Send us the device's intended purpose, its current classification and whatever technical documentation already exists — in any language. We will check it against the EAEU classification rules, identify the evidence that is missing, and give you a stage-by-stage plan with realistic dates.

There is no charge for that first assessment and it usually takes one working day. Our services and indicative pricing are on the services and tariffs pages; you can also read about the specialists who would run the project on the team page. To start, use the contacts page.

Frequently asked questions

How many reference states can I choose for one device?
Only one: for a single device you may not select two or more reference states. The authority and expert organisation of that state carry out the examination and issue the certificate.
How long does the procedure take?
The regulator's own stages add up to roughly five months: 7 working days to check that the dossier is complete, 45 working days for the examination, 30 working days for each recognition state, then 10 plus 10 working days for the decision and the certificate. A full project with testing, translation and dossier preparation normally runs 12 to 24 months.
What happens if a recognition state does not confirm the expert conclusion?
The information about that state is removed from the certificate, so the device stays registered for the remaining states only.
Do we have to use the Union procedure now, or can we still register nationally?
Applications may be submitted until 31 December 2027 either under Union law or under the national legislation of a member state, under the Protocol signed on 29 December 2025. A national certificate is valid in one country only, so other member states would require separate registration.
We are based outside the Union — what does that require of us?
A manufacturer from a third state must appoint an authorised representative resident in a member state and keep its details up to date in the dossier. Otherwise the requirements are identical to those for devices made inside the Union.